Quick summary: Endotracheal Tube, Cuffed. Reason: Health product recall. Category: Health products. Date: 2026-08-14.
Endotracheal Tube, Cuffed
2026-08-14
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Endotracheal Tube, Cuffed. Reason: Health product recall. Category: Health products. Date: 2026-08-14.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
The 6FR suction catheter could not pass smoothly through the 2.5mm sizes of cuffed tracheal tube. This may lead to reintubations if the airway is unable to be suctioned.To date, there have been two complaints reported from united states customers where one required reintubation. No serious patient harm has occurred.
What you should do
- Contact the manufacturer if you require additional information.
Additional official information