Public recall detail

EMIT® 2000 Quinidine Assay

2026-10-15

medium
Health products Health product recall
Quick summary

Quick summary: EMIT® 2000 Quinidine Assay. Reason: Health product recall. Category: Health products. Date: 2026-10-15.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: EMIT® 2000 Quinidine Assay. Reason: Health product recall. Category: Health products. Date: 2026-10-15.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandSiemens Healthcare Diagnostics Inc.
ProductEMIT® 2000 Quinidine Assay
IssueMedical devices - Performance
Recall classType II
IdentificationRA-82788

Affected products

ProductEMIT® 2000 Quinidine Assay
Lot2597
ModelOSR4Q229
Serial number2597
DescriptionEMIT® 2000 Quinidine Assay
ProductEMIT® 2000 Quinidine Assay
Lot2709
ModelOSR4Q229
Serial number2709
DescriptionEMIT® 2000 Quinidine Assay
ProductEMIT® 2000 Quinidine Assay
Lot2627
ModelOSR4Q229
Serial number2627
DescriptionEMIT® 2000 Quinidine Assay

Issue

Siemens Healthineers has confirmed through the investigation of customer complaints and internal investigation that the potential for Repeatability precision to be greater than 10% coefficient of variation (CV) and Within-Lab precision to be greater than 15% CV. Using a 20 - day ANOVA study, Repeatability ranged from 3.09% to 52.03% CV across the assay range when using multiple lots of reagents on the DXC 700 AU and the Viva-ProE. In the same study, Within-Lab precision ranged from 7.72% to 53.39% CV across the assay range.

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - Chemistry
CompaniesSiemens Healthcare Diagnostics Inc. 500 Gbc Drive, Newark, Delaware, United States, 19714
Recall date2026-09-21
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-82788
Background{}
CompaniesSiemens Healthcare Diagnostics Inc. 500 Gbc Drive, Newark, Delaware, United States, 19714
AudiencesHealthcare
Source updated2026-10-15

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.