Public recall detail

Embletta MPR PG

2026-10-06

medium
Health products Health product recall
Quick summary

Quick summary: Embletta MPR PG. Reason: Health product recall. Category: Health products. Date: 2026-10-06.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Embletta MPR PG. Reason: Health product recall. Category: Health products. Date: 2026-10-06.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandNatus Medical Incorporated Adba Excel-Tech Ltd
ProductEmbletta MPR PG
IssueMedical devices - Performance
Recall classType II
IdentificationRA-82736

Affected products

ProductEmbletta MPR PG
LotEBMPG-C1013446
Model2003011
Serial numberEBMPG-C1013446
DescriptionEmbletta MPR PG
ProductEmbletta MPR PG
LotEBMPG-C1013450
Model2003011
Serial numberEBMPG-C1013450
DescriptionEmbletta MPR PG
ProductEmbletta MPR PG
LotEBMPG-C1013674
Model2003011
Serial numberEBMPG-C1013674
DescriptionEmbletta MPR PG
ProductEmbletta MPR PG
LotEBMPG-C1013449
Model2003011
Serial numberEBMPG-C1013449
DescriptionEmbletta MPR PG
ProductEmbletta MPR PG
LotEBMPG-C1013457
Model2003011
Serial numberEBMPG-C1013457
DescriptionEmbletta MPR PG
ProductEmbletta MPR PG
LotEBMPG-C1013677
Model2003011
Serial numberEBMPG-C1013677
DescriptionEmbletta MPR PG
ProductEmbletta MPR PG
LotEBMPG-C1013435
Model2003011
Serial numberEBMPG-C1013435
DescriptionEmbletta MPR PG
ProductEmbletta MPR PG
LotEBMPG-C1013451
Model2003011
Serial numberEBMPG-C1013451
DescriptionEmbletta MPR PG

Issue

Natus has become aware of complaints related to certain Embletta MPR devices where the device is unable to correctly format its internal memory. When the internal memory cannot be correctly formatted, previously recorded studies may remain on the device and may be updated with incorrect patient information from a subsequent study. The issue may also prevent studies from initiating recording successfully or may cause recordings to end prematurely, resulting in incomplete or unavailable diagnostic data and the potential need to repeat a patient study. In some cases, users may be unable to download previously recorded studies from the device. Customers are unable to correct the memory formatting issue through normal device operation, and there is currently no workaround available. These conditions may increase the risk of incorrect patient information being associated with a study, loss of diagnostic data, delays in diagnosis, or the need to repeat patient testing.

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - General hospital and personal use
CompaniesNatus Medical Incorporated Adba Excel-Tech Ltd (Xltek) 2568 Bristol Circle, Oakville, Ontario, Canada, L6H 5S1
Recall date2026-09-24
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-82736
Background{}
CompaniesNatus Medical Incorporated Adba Excel-Tech Ltd (Xltek) 2568 Bristol Circle, Oakville, Ontario, Canada, L6H 5S1
AudiencesHealthcare
Source updated2026-10-06

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.