Quick summary: EMAG® Instrument. Reason: Health product recall. Category: Health products. Date: 2026-10-02.
EMAG® Instrument
2026-10-02
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: EMAG® Instrument. Reason: Health product recall. Category: Health products. Date: 2026-10-02.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
bioMérieux received complaints from the field related to EMAG® system Reference 418591 Error 10003 Heater timeout issues. The risk associated with the referenced issue is a potential delayed result / no result (loss of sample). Initial investigation of the issue identified the error started to occur on the field after firmware SP4 Patch 2 was installed. The investigation is ongoing to identify the root cause of the 10003-error heater timeout issue.
What you should do
- Contact the importer.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.