Elekta Modulator Discharge Rod
2025-11-19
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Elekta Modulator Discharge Rod. Date: 2025-11-19.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Elekta has identified a safety concern related to the modulator discharge rod. It was discovered that the supplier had been incorrectly assembling the crimp within the crimping tool - fitting it in reverse. As a result, the earth lead on the safety discharge rod was able to be pulled out with minimal force, which could pose potential safety risks to the field service engineer. While not all crimps are affected, reversed crimps were found in stock at Elekta.
What you should do
- Elekta will issue an IFSM to check and replace faulty discharge rod crimps as needed. A service representative will contact users.