Quick summary: Edwards Lifesciences Swan-Ganz Catheters. Reason: Performance. Category: Cardiovascular. Date: 2026-04-29.
Edwards Lifesciences Swan-Ganz Catheters
2026-04-29
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Edwards Lifesciences Swan-Ganz Catheters. Reason: Performance. Category: Cardiovascular. Date: 2026-04-29.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
There have been a growing number of complaints (outside of Canada) of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on Swan-Ganz catheters. The issue is linked to manufacturing process and material changes and is associated with situations where the (blue) proximal injectate hub and lumen experience frequent manipulation or flexing. Most complaints are associated with performance of bolus thermodilution cardiac output procedures. This has led to instances of leakage, lumen damage and breakage and may lead to infection, medication loss, and/or blood loss.
What you should do
- Contact the manufacturer if you require additional information.
Additional official information