Quick summary: Easy Moving Plus B Mobile X-Ray Unit With Digital Detector. Date: 2024-06-20.
Easy Moving Plus B Mobile X-Ray Unit With Digital Detector
2024-06-20
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Easy Moving Plus B Mobile X-Ray Unit With Digital Detector. Date: 2024-06-20.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Notification to inform customers regarding potential safety issue related to the steel cable installed inside the mobile column which supports the weight of telescopic arm with the tube head. The potential risk consists of an interference which could cause that the safety system which blocks the arm of the cable is cut off does not work properly. The interference occurs between the cable and a metal plate of the arm's carriage due to shape of the cable's crimped terminals. Scope of the safety recall is to review the installed base according to the manufacturer's instructions and install the kit immediately if the cable could interfere with the parachute system. Recall start date: June 5, 2024
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...