DxC 500i Clinical Analyzer and DxC 500i AU Chemistry Module
2026-11-18
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: DxC 500i Clinical Analyzer and DxC 500i AU Chemistry Module. Reason: Health product recall. Category: Health products. Date: 2026-11-18.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Beckman Coulter has identified an issue that may affect the reporting of clinical chemistry test results. When a laboratory selects a reporting unit other than the system regional default unit for a clinical chemistry test and multiple tests are ordered under the same sample ID, or multiple replicates of the same test under the same sample ID, the software may incorrectly apply the unit conversion factor more than once. This may result in an erroneous reported value.
What you should do
- Contact the manufacturer if you require additional information.