Quick summary: Dual Flat Panel Monitor (DFPM) Accessory To The HarmonyAIR® A-Series And E-Series Surgical Lighting Systems. Date: 2025-06-24.
Dual Flat Panel Monitor (DFPM) Accessory To The HarmonyAIR® A-Series And E-Series Surgical Lighting Systems
2025-06-24
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Dual Flat Panel Monitor (DFPM) Accessory To The HarmonyAIR® A-Series And E-Series Surgical Lighting Systems. Date: 2025-06-24.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Steris identified that the DFPM yoke assembly may detach from the spring arm. Should this occur, the yoke assembly is expected to hang by its wiring rather than fully detach, potentially requiring patient transfer and causing a procedure delay. In addition, the yoke assembly could come in contact with the user or patient during detachment. This could result in an injury. Users are requested to review revised preventive maintenance manual and interval based check list language for the HarmonyAIR® a series surgical lighting system and immediately complete maintenance activities as instructed. Recall start date: June 13, 2025
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.