Quick summary: Digital Diagnostic Mobile X-Ray System GM85. Date: 2025-10-28.
Digital Diagnostic Mobile X-Ray System GM85
2025-10-28
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Digital Diagnostic Mobile X-Ray System GM85. Date: 2025-10-28.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
We have received a report regarding the loosening or breakage of bolts that connect the tube head unit (THU) to the arm. An investigation determined that this may be caused by metal fatigue resulting from accumulated stress over a long period of use. If all connecting bolts were to fail, it could potentially cause the THU to detach from the arm, posing a risk of serious injury to a patient or user.
What you should do
- Customers to be contacted to arrange visit for inspection and repair.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.