Delfia Inducer

2023-02-16

medium
Recall Safety notice
Quick summary

Quick summary: Delfia Inducer. Date: 2023-02-16.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Delfia Inducer. Date: 2023-02-16.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandWallac Oy
ProductDelfia Inducer
IssueMedical devices - Labelling and packaging
Recall classType II
IdentificationRA-72686

Affected products

ProductDelfia Inducer
Lot742762 742802
Model3027-0010
Serial number742762 742802
DescriptionDelfia Inducer

Issue

On Delfia Inducer bottles (2 kit lots), the label is placed too low preventing the Delfia Xpress instrument level sensor to detect the Delfia inducer liquid level. Because of that the instrument does not give an 'inducer bottle soon empty' warning at level where there is 4 ml Delfia inducer left (30 tests + 1 ml dead volume). This may cause the Delfia inducer to run out and this in turn may cause failed or false results. Recall start date: February 7, 2023

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - Chemistry
CompaniesWallac Oy Mustionkatu 6, Turku, Finland, 20750
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-72686
Background{}
CompaniesWallac Oy Mustionkatu 6, Turku, Finland, 20750
AudiencesHealthcare
Source updated2023-02-16

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.