Quick summary: Datascope Cardiosave Hybrid And Rescue Intra-Aortic Balloon Pumps (IABP). Date: 2023-06-16.
Datascope Cardiosave Hybrid And Rescue Intra-Aortic Balloon Pumps (IABP)
2023-06-16
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Datascope Cardiosave Hybrid And Rescue Intra-Aortic Balloon Pumps (IABP). Date: 2023-06-16.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Issue 1: There have been reports of IABP units losing the ability to charge batteries in one or both bay slots due to a failure in the power management board charging path circuitry due to damage to the component from a previous electrical surge. Therapy may be interrupted if batteries fail to charge and the device is disconnected from ac power. Low battery alarms may alert the user to the issue prior to interruption of therapy. Issue 2: An unexpected shutdown of the IABP may occur due to failures of tantalum capacitors in the power management board and/or solenoid board. This issue may lead to an unexpected interruption of therapy. Recall start date: June 9, 2023
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.