Quick summary: CyberKnife® Treatment Delivery System. Reason: Health product recall. Category: Health products. Date: 2026-06-22.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandAccuray, Inc.
ProductCyberKnife® Treatment Delivery System
IssueMedical devices - Performance
Recall classType II
IdentificationRA-82239

Affected products

ProductCyberKnife® Treatment Delivery System
LotC0417
Model0660000
Serial numberC0417
DescriptionCyberKnife® Treatment Delivery System
ProductCyberKnife® Treatment Delivery System
LotC0545
Model0660000
Serial numberC0545
DescriptionCyberKnife® Treatment Delivery System
ProductCyberKnife® Treatment Delivery System
LotC0573
Model0660000
Serial numberC0573
DescriptionCyberKnife® Treatment Delivery System
ProductCyberKnife® Treatment Delivery System
LotC0351
Model0660000
Serial numberC0351
DescriptionCyberKnife® Treatment Delivery System

Issue

During collimator housing drop-off or pickup over the Xchange® table, the system may interlock, stop motion, and prompt service contact. Remove the patient before troubleshooting, do not move the treatment robot using the in-room pendant, and contact accuray service.

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - Radiology
CompaniesAccuray, Inc.1209 Deming Way, Madison, Wisconsin, United States, 53717
Recall date2026-06-11
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-82239
Background{}
CompaniesAccuray, Inc.1209 Deming Way, Madison, Wisconsin, United States, 53717