Quick summary: Cyanosure DermaV Laser System. Reason: Health product recall. Category: Health products. Date: 2026-08-26.
Cyanosure DermaV Laser System
2026-08-26
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Cyanosure DermaV Laser System. Reason: Health product recall. Category: Health products. Date: 2026-08-26.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Cynosure Lutronic has identified a adverse trend of complaint reports in North America for spark and loud noise resulting in patient burns during vascular treatment with the DermaV laser system. There is a remote possibility that utilizing higher fluences at 1064 nm wavelength with the current cryogen parameters can result in a spark/loud noise being observed during treatment that could potentially result in patient burn.
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...