Public recall detail

Custom Ophthalmic Kits

2024-07-03

medium
Recall Safety notice
Quick summary

Quick summary: Custom Ophthalmic Kits. Date: 2024-07-03.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Custom Ophthalmic Kits. Date: 2024-07-03.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandMicrospecialties Llc
ProductCustom Ophthalmic Kits
IssueMedical devices - Sterility
Recall classType II
IdentificationRA-75779

Affected products

ProductCustom Ophthalmic Kits
LotAll lots with the prefix P6276 & P6304
ModelM99-0061-219042C M99-0061-219031A M99-0061-219107A M99-0061-220282A M99-0061-218024B M99-0063-21437N M99-0063-21443C M99-0063-221205C M99-0063-21439J M99-0063-21492N M99-0063-218001A M99-0063-21435G M99-0063-21671G M99-0063-219094H
Serial numberAll lots with the prefix P6276 & P6304
DescriptionCustom Ophthalmic Kits

Issue

The recall is due to Microspecialties' inability to confirm that the products were exposed to the proper sterilization cycle. Recall Start Date: June 11, 2024

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - General hospital and personal use
CompaniesMicrospecialties Llc 430 Smith Street, Middletown, Connecticut, United States, 06457
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-75779
Background{}
CompaniesMicrospecialties Llc 430 Smith Street, Middletown, Connecticut, United States, 06457
AudiencesHealthcare
Source updated2024-07-03

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.

Starting...