Quick summary: Corin Operating Tables. Date: 2025-07-24.
Corin Operating Tables
2025-07-24
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Corin Operating Tables. Date: 2025-07-24.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Under certain conditions - such as the simultaneous use of the operating table with other medical devices that may cause interference, including High-Frequency (HF) surgical equipment - we have identified a potential issue that could prevent the device from performing as intended. If the issue occurs, the Corin operating table does not respond to any command. It is impossible to actuate any movement of the Corin operating table, neither with the universal remote control nor the override panel. We are not aware of instances where a patient was injured due to the described issue. Recall start date: July 18, 2025
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.