Public recall detail

Corin Operating Tables

2025-07-24

medium
Recall Safety notice
Quick summary

Quick summary: Corin Operating Tables. Date: 2025-07-24.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Corin Operating Tables. Date: 2025-07-24.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandMaquet GmbH
ProductCorin Operating Tables
IssueMedical devices - Performance
Recall classType II
IdentificationRA-77760

Affected products

ProductCorin Operating Tables
Lot7
Model770001F0
Serial number7
DescriptionCorin Operating Tables
ProductCorin Operating Tables
Lot9
Model770001F0
Serial number9
DescriptionCorin Operating Tables
ProductCorin Operating Tables
Lot13
Model770001F0
Serial number13
DescriptionCorin Operating Tables
ProductCorin Operating Tables
Lot11
Model770001F0
Serial number11
DescriptionCorin Operating Tables

Issue

Under certain conditions - such as the simultaneous use of the operating table with other medical devices that may cause interference, including High-Frequency (HF) surgical equipment - we have identified a potential issue that could prevent the device from performing as intended. If the issue occurs, the Corin operating table does not respond to any command. It is impossible to actuate any movement of the Corin operating table, neither with the universal remote control nor the override panel. We are not aware of instances where a patient was injured due to the described issue. Recall start date: July 18, 2025

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - General and plastic surgery
CompaniesMaquet Gmbh Kehler Strasse 31, Rastatt, Germany, 76437
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-77760
Background{}
CompaniesMaquet Gmbh Kehler Strasse 31, Rastatt, Germany, 76437
AudiencesHealthcare
Source updated2025-07-24

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.

Starting...

Related recalls

Similar recalls and safety notices.

Loading related recalls...