Public recall detail

Cook Medical Beacon® Tip Centimeter Sizing Catheter

2026-04-28

medium
Cardiovascular Performance
Quick summary

Quick summary: Cook Medical Beacon® Tip Centimeter Sizing Catheter. Reason: Performance. Category: Cardiovascular. Date: 2026-04-28.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Cook Medical Beacon® Tip Centimeter Sizing Catheter. Reason: Performance. Category: Cardiovascular. Date: 2026-04-28.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandCook, Incorporated
ProductCook Medical Beacon® Tip Centimeter Sizing Catheter
IssueMedical devices - Performance
Recall classType II
IdentificationRA-81989

Affected products

ProductBeacon® Tip Centimeter Sizing Catheter
LotMore than 20 numbers, contact manufacturer.
ModelN5.0-35-100-P-10S-0-CSC-20
Serial numberMore than 20 numbers, contact manufacturer.
DescriptionBeacon® Tip Centimeter Sizing Catheter
ProductBeacon® Tip Centimeter Sizing Catheter
LotMore than 20 numbers, contact manufacturer.
ModelNR5.0-35-70-P-10S-0-CSC-20
Serial numberMore than 20 numbers, contact manufacturer.
DescriptionBeacon® Tip Centimeter Sizing Catheter
ProductBeacon® Tip Centimeter Sizing Catheter
LotMore than 20 numbers, contact manufacturer.
ModelN5.0-35-100-P-10S-PIG-CSC-20
Serial numberMore than 20 numbers, contact manufacturer.
DescriptionBeacon® Tip Centimeter Sizing Catheter

Issue

Cook Medical identified that the marker bands on impacted products may be at an increased risk of cracking/breakage. The subject marker bands are provided to Cook by an external supplier. This issue was identified internally during production. Following issue detection, product samples underwent tensile testing, where one sample cracked while being manipulated to load into the test fixture. While there is a quality control check in place to detect this failure mode prior to product release, all tested samples had underwent this quality control check prior to being tested. This is indicative that the subject marker bands may fail at a point past the quality control check. This issue has been isolated to specific supplier lots of these marker bands.

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - Cardiovascular
CompaniesCook, Incorporated 750 Daniels Way, Bloomington, Indiana, United States, 47404
Recall date2026-04-02
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-81989
Background{}
CompaniesCook, Incorporated 750 Daniels Way, Bloomington, Indiana, United States, 47404
AudiencesHealthcare
Source updated2026-04-28