Quick summary: Cook Medical Beacon® Tip Centimeter Sizing Catheter. Reason: Performance. Category: Cardiovascular. Date: 2026-04-28.
Cook Medical Beacon® Tip Centimeter Sizing Catheter
2026-04-28
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Cook Medical Beacon® Tip Centimeter Sizing Catheter. Reason: Performance. Category: Cardiovascular. Date: 2026-04-28.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Cook Medical identified that the marker bands on impacted products may be at an increased risk of cracking/breakage. The subject marker bands are provided to Cook by an external supplier. This issue was identified internally during production. Following issue detection, product samples underwent tensile testing, where one sample cracked while being manipulated to load into the test fixture. While there is a quality control check in place to detect this failure mode prior to product release, all tested samples had underwent this quality control check prior to being tested. This is indicative that the subject marker bands may fail at a point past the quality control check. This issue has been isolated to specific supplier lots of these marker bands.
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.