Quick summary: cobas® c 303 analytical unit. Reason: Health product recall. Category: Health products. Date: 2026-10-15.
cobas® c 303 analytical unit
2026-10-15
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: cobas® c 303 analytical unit. Reason: Health product recall. Category: Health products. Date: 2026-10-15.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
An issue has been identified where vacuum tubing of the cell rinse unit of the cobas c 303 analytical unit prematurely deteriorated. Investigations confirmed the issue. In the event of small pinholes or cracks, no impact on analytical performance was identified. However, in case of tubing rupture, the following can occur as a result of the issue: Undetected leakage: Fluid may drip into random reaction cells. Result impact: Increased risk of erroneous patient and QC results.
What you should do
- Roche Care Center at 1-877-273-3433, via navify®/Online Support and also by e-mail at laval.techinfo@roche.com.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.