Public recall detail

Cholestech Ldx Total, HDL, Triglycerides Cholesterol and Glucose Test

medium
Recall Safety notice
Quick summary

Quick summary: Cholestech Ldx Total, HDL, Triglycerides Cholesterol and Glucose Test.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Cholestech Ldx Total, HDL, Triglycerides Cholesterol and Glucose Test.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandAlere San Diego, Inc.
ProductCholestech Ldx Total, HDL, Triglycerides Cholesterol and Glucose Test
IssueMedical devices - Labelling and packaging Medical devices - Performance
Recall classType II
IdentificationRA-64285

Affected products

ProductCholestech Ldx Total Cholesterol Test
LotAll lots.
Model10-987 10-991 10-990 10-989
Serial numberAll lots.
DescriptionCholestech Ldx Total Cholesterol Test
ProductCholestech Ldx Hdl Cholesterol Test
LotAll lots.
Model10-987 10-991 10-990 10-989
Serial numberAll lots.
DescriptionCholestech Ldx Hdl Cholesterol Test
ProductCholestech Ldx Triglycerides Test
LotAll lots.
Model10-991 10-989
Serial numberAll lots.
DescriptionCholestech Ldx Triglycerides Test
ProductCholestech Ldx Glucose Test
LotAll lots.
Model10-991 10-990
Serial numberAll lots.
DescriptionCholestech Ldx Glucose Test

Issue

Alere San Diego has identified that bilirubin interference level listed in the package insert is incorrect. Concentrations less than 5.0 mg/dl can impact the accuracy of triglyceride (TRG) and High-density lipoproteins (HDL). Bilirubin levels of 2.0 mg/dl (conjugated) or 1.6 mg/dl(unconjugated) were found to be the correct concentrations. this change impacts all lots produced since product launch. Recall start date: June 13, 2022

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare Industry
CategoryHealth products - Medical devices - Chemistry
CompaniesAlere San Diego, Inc. 9975 Summers Ridge Road, San Diego, California, United States, 92121
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-64285
Background{}
CompaniesAlere San Diego, Inc. 9975 Summers Ridge Road, San Diego, California, United States, 92121
AudiencesHealthcare Industry
Source updated2022-06-20

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.

Starting...

Related recalls

Similar recalls and safety notices.