Quick summary: Catalyst Hd and Catalyst+ Hd Systems. Date: 2025-10-21.
Catalyst Hd and Catalyst+ Hd Systems
2025-10-21
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Catalyst Hd and Catalyst+ Hd Systems. Date: 2025-10-21.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
C-RAD has noted a potential issue affecting the CSRs mode of the Catalyst+ HD and Catalyst HD systems. The issue is that the stated precision of the catalyst system for stereotactic radiosurgery (SRS) treatments cannot be guaranteed for all couch angles and system setups. in certain cases, the system may indicate the patient as correctly positioned at the isocenter, even when the isocenter is positioned outside of the indicated tolerance. This could in the worstcase result in erroneous dose delivery.
What you should do
- Contact the manufacturer if you require additional information.