Quick summary: CareLink™ SmartSync™ Device Manager. Reason: Health product recall. Category: Health products. Date: 2026-10-01.
CareLink™ SmartSync™ Device Manager
2026-10-01
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: CareLink™ SmartSync™ Device Manager. Reason: Health product recall. Category: Health products. Date: 2026-10-01.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
A complaint from the CareLink network reported a sudden decrease in the Remaining Longevity of a patient's Micra VR CIED from 3 years to 6 months following a remote transmission, which was later found to be incorrect. Analysis revealed a calculation issue in SmartSync and CareLink for Micra VR2 and Micra AV2 devices, occurring under specific conditions. This miscalculation stems from a software error where a function received an integer instead of a real data type, defaulting the average pacing rate to 60 BPM and resulting in inaccurate Remaining Longevity estimates.
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.