Quick summary: Cardiosave System - Hybrid & Rescue Intra-Aortic Balloon Pump. Date: 2023-01-04.
Cardiosave System - Hybrid & Rescue Intra-Aortic Balloon Pump
2023-01-04
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Cardiosave System - Hybrid & Rescue Intra-Aortic Balloon Pump. Date: 2023-01-04.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Datascope/Getinge has received complaints reporting unexpected shutdown of the Cardiosave IABP while providing therapy in very rare instances. An internal investigation of the complaints determined an unexpected shutdown may be due to blood entering into the Cardiosave IABP when therapy is provided with a perforated intra-aortic balloon catheter. Recall start date: December 21, 2022
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.