Quick summary: CardioMEMS Patient Electronics System. Date: 2024-10-18.
CardioMEMS Patient Electronics System
2024-10-18
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: CardioMEMS Patient Electronics System. Date: 2024-10-18.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Abbott has received reports of an issue following migration of the merlin.net patient care network hf portal to the cloud environment on September 30, 2024. Since then, certain patients have been impacted, and their patient electronics system may be sending incorrect pressure readings. While the issue is being investigated, Abbott is advising users to pause using these patients' pulmonary artery pressure readings for clinical decision-making or patient management. Recall start date: October 9, 2024
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.