Quick summary: Cardiohelp-I. Date: 2024-04-08.
Cardiohelp-I
2024-04-08
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Cardiohelp-I. Date: 2024-04-08.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
During service measurements on the Cardiohelp-I devices in the field, it was noticed that the measured patient leakage current exceeded the limit values of 50 μa specified in standard EN 60601-1:2006 + a1:2013. Patient leakage currents of up to 1337 μa were measured. Investigations have shown that only Cardiohelp-I devices with installed sensor panels (701051308) with an old sensor bridge (701049193) hardware version are affected. The new hardware version is not affected. Recall Start Date: March 18, 2024
What you should do
- Contact the manufacturer if you require additional information.