Quick summary: CADD-Solis Ambulatory Infusion Pump - Battery Pack. Date: 2024-08-20.
CADD-Solis Ambulatory Infusion Pump - Battery Pack
2024-08-20
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: CADD-Solis Ambulatory Infusion Pump - Battery Pack. Date: 2024-08-20.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Smiths Medical identified three (3) reports in which damage to the battery pack may have caused a short to a capacitor within the battery pack. While the battery encasement is designed to be flame retardant, a short to the capacitor may potentially lead to melting of the battery pack case. If this issue occurs, the battery pack charging circuit may become inoperable. Recall start date: July 22, 2024
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.