CADD-Solis Ambulatory Infusion Pump
2025-04-23
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: CADD-Solis Ambulatory Infusion Pump. Date: 2025-04-23.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
CADD-Solis pumps may lose communication with the CADD wireless communication module if wireless network setting changes on the hospital network are not compatible with the CADD wireless communication module. This may result in a "wireless module intermittent connection" high priority alarm, which will stop an ongoing infusion. Recall Start Date: April 8, 2025
What you should do
- Contact the manufacturer if you require additional information.