Public recall detail

Cadd Administration Set

2022-12-21

medium
Recall Safety notice
Quick summary

Quick summary: Cadd Administration Set. Date: 2022-12-21.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Cadd Administration Set. Date: 2022-12-21.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandSmiths Medical Asd, Inc.
ProductCadd Administration Set
IssueMedical devices - Performance
Recall classType II
IdentificationRA-72397

Affected products

ProductMedication Cassett Reservior
LotMore than 10 numbers, contact manufacturer.
Model21-7308-24 21-7301-24 21-7309-24 21-7300-24 21-7310-24 21-7302-24
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionMedication Cassett Reservior
ProductCadd Yellow Administration Set With Nrfit Connectors
LotMore than 10 numbers, contact manufacturer.
Model21-7624-24 21-7649-24
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionCadd Yellow Administration Set With Nrfit Connectors
ProductCadd Administration Set
LotMore than 10 numbers, contact manufacturer.
Model21-7321-24 21-7324-24 21-7349-24 21-7395-24 21-7339-24 21-7343-24 21-7391-24 21-7333-24 21-7394-24 21-7346-24 21-7323-24 21-7336-24 21-7363-24 21-7383-24 21-7390-24 21-7322-24 21-7359-24
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionCadd Administration Set

Issue

Manufacturing variations may cause the green cadd flow stop arm to compress and partially occlude the tubing before clinical use. If this occurs, there is a potential that the occlusion does not resolve when an affected reservoir or administration set is connected to the pump, and the pump may not detect the occlusion. This may result in underdelivery, despite the pump displaying that the infusion is running properly. Cadd flow stop medication cassette reservoirs and cadd flow stop administration sets provide free-flow protection. by design, a green, spring-loaded pivoting arm automatically squeezes the tubing closed when the reservoir or administration set is not installed on a pump. Connecting the reservoir or administration set to the pump causes the pump to push the flow stop arm, enabling fluid flow through the tubing. In certain circumstances, the tubing may remain occluded even though the cadd reservoir or administration set is loaded into the pump. Recall start date: November 29, 2022

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - General hospital and personal use
CompaniesSmiths Medical Asd, Inc. 6000 Nathan Lane N., Minneapolis, Minnesota, United States, 55442
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-72397
Background{}
CompaniesSmiths Medical Asd, Inc. 6000 Nathan Lane N., Minneapolis, Minnesota, United States, 55442
AudiencesHealthcare
Source updated2022-12-21

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.