Public recall detail

Blanketrol® II 115V and Blanketrol® III 115V

2026-07-28

medium
Health products Health product recall
Quick summary

Quick summary: Blanketrol® II 115V and Blanketrol® III 115V. Reason: Health product recall. Category: Health products. Date: 2026-07-28.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Blanketrol® II 115V and Blanketrol® III 115V. Reason: Health product recall. Category: Health products. Date: 2026-07-28.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandGentherm Medical, LLC
ProductBlanketrol® II 115V and Blanketrol® III 115V
IssueMedical devices - Performance
Recall classType II
IdentificationRA-82413

Affected products

ProductBlanketrol® II 115V
Lot151-2-01474
Model222S
Serial number151-2-01474
DescriptionBlanketrol® II 115V
ProductBlanketrol® II 115V
Lot151-2-01472
Model222S
Serial number151-2-01472
DescriptionBlanketrol® II 115V
ProductBlanketrol® III 115V
LotMore than 10 numbers, contact manufacturer.
Model233
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionBlanketrol® III 115V

Issue

Gentherm Medical has received complaints of overheated circuit boards in 0.53% of distributed devices. Overheating can cause the circuit board to stop operating, which may result in a device that shuts down at startup or during use on a patient.Based on the low probability of harm, and absence of reported harm, associated with this issue, users may continue to use the device. In an abundance of caution, it is recommended to have an alternative means of providing therapy available.

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare Industry
CategoryHealth products - Medical devices - General hospital and personal use Health products - Medical devices - Cardiovascular
CompaniesGentherm Medical, LLC
Recall date2026-07-13
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-82413
Background{}
CompaniesGentherm Medical, LLC
AudiencesHealthcare Industry
Source updated2026-07-28

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.

Starting...

Related recalls

Similar recalls and safety notices.