Quick summary: BIOFIRE® Filmarray® Gastrointestinal (GI) Panel. Date: 2024-07-18.
BIOFIRE® Filmarray® Gastrointestinal (GI) Panel
2024-07-18
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: BIOFIRE® Filmarray® Gastrointestinal (GI) Panel. Date: 2024-07-18.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
An increased number of customer complaints associated with Vibrio/Vibrio Cholerae false positive detections by the BIOFIRE Filmarray Gastrointestinal (GI) Panel have been observed when using Remel Cary-Blair transport medium. The BIOFIRE GI panel and the Cary-Blair transport medium are performing as intended. Per INV-19261, the root cause of the false positive Vibrio/Vibrio Cholerae results detected by the BIOFIRE GI panel is the presence of non-viable organism in the Remel Cary-Blair transport medium. Recall Start Date: July 4, 2024
What you should do
- Contact the importer, Biomérieux Canada Inc.