Quick summary: Beckman Coulter AU Immunochemistry System - AU480 Analyzer. Reason: Health product recall. Category: Health products. Date: 2026-08-06.
Beckman Coulter AU Immunochemistry System - AU480 Analyzer
2026-08-06
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Beckman Coulter AU Immunochemistry System - AU480 Analyzer. Reason: Health product recall. Category: Health products. Date: 2026-08-06.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Beckman Coulter has determined that a software defect in AU480 chemistry analyzers using Microsoft® SQL server 2022 express may cause a DB connection error that renders alarmDB unreadable and prevents instrument startup. The issue is associated with transaction log auto-growth, and may cause delays in reporting patient test results to physicians.
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.