Affected products
Issue
UIBC OSR61205 is not meeting interference claims for haemoglobin interference serum application. As per product ifus the claim is less than 10% up to 2g/l haemoglobin. While the high pool is meeting the claim, the low pool is failing the hemolysate interference specifications with results of -43.63% or -8.95 μmol/l. Recall start date: July 1, 2025
What you should do
- Contact the manufacturer if you require additional information.
Additional official information