Public recall detail

Becker and Exacta External Drainage and Monitoring Systems

2024-11-14

medium
Recall Safety notice
Quick summary

Quick summary: Becker and Exacta External Drainage and Monitoring Systems. Date: 2024-11-14.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Becker and Exacta External Drainage and Monitoring Systems. Date: 2024-11-14.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandMEDTRONIC INC.
ProductBecker and Exacta External Drainage and Monitoring Systems
IssueMedical devices - Performance
Recall classType II
IdentificationRA-76501

Affected products

ProductExacta With Edm Ventricular Catheter Kit (Barium)
LotMore than 10 numbers, contact manufacturer.
Model27581
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionExacta With Edm Ventricular Catheter Kit (Barium)
ProductBecker External Drainage And Monitoring System With Blue Core Needleless Injection Site And Bonded Patient Line Connections
LotMore than 10 numbers, contact manufacturer.
Model27779
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionBecker External Drainage And Monitoring System With Blue Core Needleless Injection Site And Bonded Patient Line Connections
ProductBecker External Drainage And Monitoring System
LotMore than 10 numbers, contact manufacturer.
Model46128
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionBecker External Drainage And Monitoring System
ProductExacta External Drainage And Monitoring System With Blue Core Needleless Injection Sites
LotMore than 10 numbers, contact manufacturer.
Model27785
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionExacta External Drainage And Monitoring System With Blue Core Needleless Injection Sites
ProductExacta Drainage System, 100ml, With Luer Lock
LotMore than 10 numbers, contact manufacturer.
Model46705
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionExacta Drainage System, 100ml, With Luer Lock
ProductExacta External Drainage And Monitoring System
LotMore than 10 numbers, contact manufacturer.
Model46700
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionExacta External Drainage And Monitoring System

Issue

Medtronic has received customer complaints reporting cracks and/or leaks on the stopcock of Becker and Exacta EDMS devices. On these devices, the stopcocks may be in three different locations in the EDMS, depending on the specific system configuration. Cracks or leaks of the stopcock can compromise line integrity and, as a result, present a potential for infection. Three adverse events associated with patient infection have been reported. Recall start date: November 1, 2024

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - Neurology
CompaniesMedtronic Inc. 710 Medtronic Parkway N.E., Minneapolis, Minnesota, United States, 55432
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-76501
Background{}