Quick summary: BD Synapsys™ Informatics Solution. Date: 2024-09-26.
BD Synapsys™ Informatics Solution
2024-09-26
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: BD Synapsys™ Informatics Solution. Date: 2024-09-26.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
BD identified through one (1) complaint that for BD Kiestra™ customers with BD Synapsys™, running Synapsys™ software versions 4.20, 4.30, 5.10, 5.20, or 5.30 and utilizing the in-line breakpoint feature, an error is occurring where BD Synapsys™ will incorrectly display breakpoints for organisms that do not have defined breakpoints in CLSI or EUCAST. This occurs when the identified organism is included in the same interpretive group as a species for which there is a specific breakpoint. Recall start date: September 12, 2024
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.