BD Facsduet™ Sample Preparation Systems

2024-08-27

medium
Recall Safety notice
Quick summary

Quick summary: BD Facsduet™ Sample Preparation Systems. Date: 2024-08-27.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: BD Facsduet™ Sample Preparation Systems. Date: 2024-08-27.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandBecton, Dickinson And Company Bd Biosciences
ProductBD Facsduet™ Sample Preparation Systems
IssueMedical devices - Performance
Recall classType II
IdentificationRA-75970

Affected products

ProductBd Facsduet™ Sample Preparation System (Base)
Lot9220010275 9220010313
Model662588
Serial number9220010275 9220010313
DescriptionBd Facsduet™ Sample Preparation System (Base)
ProductBd Facsduet™ Premium Sample Preparation System
Lot9230000104 9230000114 9230000112 9230000105 9230000138 9230000106
Model666339
Serial number9230000104 9230000114 9230000112 9230000105 9230000138 9230000106
DescriptionBd Facsduet™ Premium Sample Preparation System

Issue

Through a customer complaint investigation, BD identified a specimen volume bias issue with the BD Facsduet™ sample preparation system when using software version 1.4.1. When the multidispense feature is enabled in the BD Facsduet™ sample preparation system software, and the user is preparing a two-tube assay, the specimen volume dispensed into the first tube will be at least 3% higher than the specimen volume dispensed into the second tube. For assays, where absolute counts are calculated using BD Trucount™ tubes, an increase in the volume may result in higher-than-expected absolute counts in the first tube compared to the second tube. This issue will be observed when preparing the following assays on the BD Facsduet™ sample preparation system: 1. BD Multitest™ IMK kit with BD Trucount™ tubes. note: if using the BD Multitest™ IMK kit with BD Trucount™ tubes, a QC message will appear on your report if the bias is outside the acceptable range per assay requirements. 2. BD Multitest™ 6-color TBNK with BD Trucount™ tubes only if consecutive replicates of two are run. 3. for all other quantitative assays utilizing BD Trucount™ tubes in consecutive replicates of two, please follow your institution's procedures to determine if a review of reported results is necessary. This issue does not affect assays that are determining percent positive, semi quantitative information. Recall start date: August 12, 2024

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - Chemistry
CompaniesBecton, Dickinson And Company Bd Biosciences 155 North Mccarthy Boulevard, Milpitas, California, United States, 95035
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-75970
Background{}
CompaniesBecton, Dickinson And Company Bd Biosciences 155 North Mccarthy Boulevard, Milpitas, California, United States, 95035
AudiencesHealthcare
Source updated2024-08-27

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.