Quick summary: BD Alaris™ EtCO2, PCA, Syringe, and Pump Modules. Date: 2024-08-30.
BD Alaris™ EtCO2, PCA, Syringe, and Pump Modules
2024-08-30
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: BD Alaris™ EtCO2, PCA, Syringe, and Pump Modules. Date: 2024-08-30.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
BD has identified an increase in bent contacts on new Female/Left IUI connectors on devices and spare parts manufactured between Jan 01, 2023 and June 06, 2024. BD was made aware of this issue through complaints identified during the device check-in process and has not received any reports of this issue occurring during patient use. A defective pin on certain Female/Left IUI connectors that are not detected prior to use could result in an inoperable device or an operable device where adjoining device cannot be powered on at start up. Recall Start Date: August 15, 2024