Baxter Corporation’s MiniCap Extended Life Peritoneal Dialysis Transfer Sets: Certain cleaning products, including hand sanitizers and solvents, may damage sets

2022-02-16

medium
Recall Safety notice
Quick summary

Quick summary: Baxter Corporation’s MiniCap Extended Life Peritoneal Dialysis Transfer Sets: Certain cleaning products, including hand sanitizers and solvents, may damage sets. Date: 2022-02-16.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Baxter Corporation’s MiniCap Extended Life Peritoneal Dialysis Transfer Sets: Certain cleaning products, including hand sanitizers and solvents, may damage sets. Date: 2022-02-16.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

Brand
ProductMiniCap Extended Life Peritoneal Dialysis Transfer Sets manufactured by Baxter Corporation (see list of affected products)
IssueMedical devices
IdentificationRA-63932

Affected products

ProductMinicap Extended Life PD Transfer Set With Roller Clamp (15cm)
Codes5C4449
ProductMinicap Extended Life PD Transfer Set With Twist Clamp (15cm)
Codes5C4482
ProductMinicap Extended Life PD Transfer Set With Twist Clamp (Extra Long) (22cm)
CodesR5C4484
ProductMinicap Extended Life PD Transfer Set With Twist Clamp (Extra Short) (10cm)
Codes5C4483

Issue

Health Canada is advising patients who use MiniCap Extended Life Peritoneal Dialysis Transfer Sets, manufactured by Baxter Corporation, that direct contact between the sets and certain cleaning products, including hand sanitizers and solvents, may damage the sets. Damaged sets could leak or crack, which may cause microbial contamination leading to the development of peritonitis in some cases. Peritonitis is the inflammation of the lining of the abdomen, caused by infection, which can be serious if not promptly treated. Symptoms include nausea, vomiting, and abdominal pain or tenderness. If you experience symptoms of peritonitis, seek medical attention immediately. Damaged sets may also lead to a delay in therapy as you seek medical assistance to change your transfer set. A transfer set is tubing that connects the catheter in the body to the bag of dialysis solution in the treatment of renal (kidney) failure. Using a transfer set helps protect the catheter from damage that may be caused by frequent clamping during dialysis exchange procedures. These transfer sets are installed by trained health care providers for use in health care and home settings. Peritoneal dialysis patients receiving dialysis at home need to open and close the transfer set several times a day to connect to fresh dialysis solution and drain existing solution from their abdomen. Baxter Corporation has informed Health Canada that, to date, it has received 13 reports internationally of peritonitis possibly related to this issue, including one Canadian case. The company has also informed Health Canada that it has notified all affected customers (health care facilities), and will update the product Instructions for Use to warn that use of the following products should be avoided on the transfer sets: hand sanitizers and cleaning products that contain hydrogen peroxide, bleach, alcohol or antiseptic agents; and solvents intended to remove adhesive residue, such as those containing acetone, toluene, xylene, or cyclohexanone The Department is monitoring the effectiveness of this corrective action and will inform Canadians if any new health risks are identified.

What you should do

  • Avoid direct contact between the transfer sets and the following cleaning products: hand sanitizers and cleaning products that contain hydrogen peroxide, bleach, alcohol or antiseptic agents; and solvents intended to remove adhesive residue, such as those containing acetone, toluene, xylene or cyclohexanone.
  • hand sanitizers and cleaning products that contain hydrogen peroxide, bleach, alcohol or antiseptic agents; and
  • solvents intended to remove adhesive residue, such as those containing acetone, toluene, xylene or cyclohexanone.
  • If your transfer set is damaged (e.g., cracked or leaking), contact your health care provider immediately for a replacement. Consult your health care provider with any questions or concerns about your transfer set.
  • Seek immediate medical attention if you experience symptoms of peritonitis. Symptoms include nausea, vomiting, and abdominal pain or tenderness.
  • Use proper cleaning practices to clean the device, as outlined in the Instructions for Use or as directed by a health care professional.
  • You may also contact Baxter Corporation at 1-866-968-7477 if you have questions about this issue or about your transfer set.
  • Report any health product-related complaints to Health Canada.
Additional official information
AudienceGeneral public
CategoryHealth products - Medical devices
CompaniesPublished by Health Canada
Published byHealth Canada
Alert / recall typePublic advisory
Identification numberRA-63932
Background{}
CompaniesPublished by Health Canada
AudiencesGeneral public
Source updated2022-02-16

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.