Quick summary: Bard Marquee 12g X 10 cm. Date: 2026-01-22.
Bard Marquee 12g X 10 cm
2026-01-22
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Bard Marquee 12g X 10 cm. Date: 2026-01-22.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
In September 2024, BD initiated a field action for specific lots of the Marquee™ Disposable Core Biopsy Instruments that were potentially manufactured missing the backend notch retention feature on the needle (figure 1), which results in inadequate securing of the needle inside the device. As part of the corrective actions associated with that issue, BD inspected all remaining needle sub-assembly lots that remained in stock at the manufacturing location. BD has since identified that one sub-assembly lot that underwent this inspection may not have been adequately sorted, thus there is a potential for specific lots of the Marquee™ Disposable Core Biopsy Instruments that were manufactured with that sub-assembly to contain needles that were not adequately inspected for the presence of the backend notch.
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.