Public recall detail

B. Braun Hemodialysis Blood Tubing Set

2026-04-01

medium
Cardiovascular Performance
Quick summary

Quick summary: B. Braun Hemodialysis Blood Tubing Set. Reason: Performance. Category: Cardiovascular. Date: 2026-04-01.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: B. Braun Hemodialysis Blood Tubing Set. Reason: Performance. Category: Cardiovascular. Date: 2026-04-01.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandB. Braun Medical Inc.
ProductB. Braun Hemodialysis Blood Tubing Set
IssueMedical devices - Performance
Recall classType I
IdentificationRA-81821

Affected products

ProductHemodialysis Blood Tubing Set
LotAll lots.
ModelB3-4630M4705
Serial numberAll lots.
DescriptionHemodialysis Blood Tubing Set
ProductHemodialysis Blood Tubing Set
LotAll lots.
ModelB3-3632M3705
Serial numberAll lots.
DescriptionHemodialysis Blood Tubing Set

Issue

The potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - Cardiovascular
CompaniesB. Braun Medical Inc. 824 12th Avenue, Bethlehem, Pennsylvania, United States, 18018
Recall date2026-03-20
Published byHealth Canada
Recall classType I
Alert / recall typeHealth product recall
Identification numberRA-81821
Background{}
CompaniesB. Braun Medical Inc. 824 12th Avenue, Bethlehem, Pennsylvania, United States, 18018
AudiencesHealthcare
Source updated2026-05-13

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.

Starting...

Related recalls

Similar recalls and safety notices.