Quick summary: Azurion and Allura Xper Systems. Reason: Performance. Category: Radiology. Date: 2026-05-06.
Azurion and Allura Xper Systems
2026-05-06
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Azurion and Allura Xper Systems. Reason: Performance. Category: Radiology. Date: 2026-05-06.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the x-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message "low load fluoroscopy flavor selected: tube cooler problem". This condition cannot be resolved by restarting the system. Under normal conditions, any leaked oil is contained within the system. In rare cases of more significant leakage, oil may extend beyond the intended containment area. If oil then comes into contact with hot surfaces, fumes may be generated. these fumes may activate environmental detection systems.
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.