Azurion 7 M20
2024-01-04
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Azurion 7 M20. Date: 2024-01-04.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Philips has become aware of a potential safety issue with the Azurion 7 M20 Flexarm System ("flexarm") with a Maquet Magnus Table with software versions 05.02.12 and 05.02.15 ("table"). Due to a communication issue between the flexarm and the table, motorized movements of the table and/or the flexarm may be lost. A representative of Getinge (manufacturer of the Maquet Magnus Table) will arrange a site visit to return the software of your table to version 05.02.12. After this change, your system will be subject to the medical device recall letter (HS23039R/2023-IGT-BST-020). Recall start date: December 13, 2023
What you should do
- Contact the manufacturer if you require additional information.