Quick summary: Avanta Fluid Injection System - Multi-Patient Sterile Disposable Sets. Reason: Performance. Category: Cardiovascular. Date: 2026-05-14.
Avanta Fluid Injection System - Multi-Patient Sterile Disposable Sets
2026-05-14
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Avanta Fluid Injection System - Multi-Patient Sterile Disposable Sets. Reason: Performance. Category: Cardiovascular. Date: 2026-05-14.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
In April 2025, an off-the-shelf stopcock used in the Avanta Multi-Patient Administration Tube sets (AVA 500 MPAT (part 2 of 2)) was revised by the supplier to achieve ISO 80369-7 compliance and ensure connectability to other connectors. These changes resulted in a fit issue with the CFCV interface on the Avanta injector system. Multiple customer reports indicate that the stopcock cannot be reliably engaged with the Contrast Flow Control Valve (CFCV) snap interface. In instances where engagement is achieved through applied force, the connection is unstable, and the stopcock may disengage before or during contrast injection, leading to procedure interruption.
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.