Automated Impella Controller (AIC)

2026-08-27

medium
Health products Health product recall
Quick summary

Quick summary: Automated Impella Controller (AIC). Reason: Health product recall. Category: Health products. Date: 2026-08-27.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Automated Impella Controller (AIC). Reason: Health product recall. Category: Health products. Date: 2026-08-27.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandAbiomed Inc.
ProductAutomated Impella Controller (AIC)
IssueMedical devices - Performance
Recall classType II
IdentificationRA-82534

Affected products

ProductAutomated Impella Controller (AIC)
LotAll lots.
Model0042-0040-CA
Serial numberAll lots.
DescriptionAutomated Impella Controller (AIC)
ProductAutomated Impella Controller (AIC)
LotAll lots.
Model0042-0000-US
Serial numberAll lots.
DescriptionAutomated Impella Controller (AIC)
ProductAutomated Impella Controller (AIC)
LotAll lots.
Model0042-0000-CA
Serial numberAll lots.
DescriptionAutomated Impella Controller (AIC)

Issue

Abiomed has initiated a voluntary medical device recall (removal) after identification of an increase in complaints related to purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly. The cause of the purge flag failures is a combination of dextrose buildup from a purge cassette leak and material embrittlement from usage, aging, or exposure to cleaning chemicals. This issue may occur from usage and is not a manufacturing defect. The purge flag component is present in all manufactured Automated Impella Controller (AIC) in Canada. However, product may continue to be used to ensure continuity of care and a prioritized risk-based strategy for component replacement will be implemented. The purge flag is the component that engages with the purge disc and allows the AIC to recognize the purge disc. These AIC purge flag component failures may result most frequently in user inconvenience due to troubleshooting, attempted purge cassette exchange and AIC exchange. If the failure occurs before initiation of support, exposure may result in a delay to procedure, which is expected to remain clinically manageable in most cases. In the case of an urgent need for mechanical circulatory support or if the failure occurs during active Impella support, exposure may result in a temporary period of inadequate hemodynamic support during controller transition.

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - Cardiovascular
CompaniesAbiomed Inc. 22 Cherry Hill Drive, Danvers, Massachusetts, United States, 01923
Recall date2026-08-18
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-82534
Background{}
CompaniesAbiomed Inc. 22 Cherry Hill Drive, Danvers, Massachusetts, United States, 01923
AudiencesHealthcare
Source updated2026-08-27