Quick summary: Automated Impella Controller. Reason: Performance. Category: Cardiovascular. Date: 2026-02-25.
Automated Impella Controller
2026-02-25
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Automated Impella Controller. Reason: Performance. Category: Cardiovascular. Date: 2026-02-25.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Abiomed has identified that Automated Impella Controller (AICS) with software V8.2.3, when used in conjunction with an Impella 5.5 with smart assist set (product code 0550-0004), has a delayed "purge system blocked" alarm. Software versions post-v10.1 have this issue addressed. Abiomed recommends that you do not use the AIC with an Impella 5.5 with smart assist set (product code 0550-0004) unless AIC software is updated to v10.1 or above.
What you should do
- Contact the manufacturer if you require additional information.