Automated Blood Coagulation Analyzers
2026-11-06
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Automated Blood Coagulation Analyzers. Reason: Health product recall. Category: Health products. Date: 2026-11-06.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
In October 2026, the legal manufacturer Sysmex notified Siemens Healthineers of an urgent field safety notice for the analyzer CN series systems.Sysmex identified that deposits may accumulate on the sample arm B probe of the CN-series analyzers.The deposits may affect sample dispensing, potentially resulting in the delivery of a lower sample volume to the reaction cuvette than defined, which and may cause falsely decreased results. Based on the manufacturer’s investigation, the accumulation of deposits occurred due to insufficient washing of the sample arm B probe independent from the assays used.
What you should do
- Contact the manufacturer if you require additional information.