Public recall detail

ARTIS Icono Ceiling, Floor and Biplane Systems

2026-06-03

medium
Health products Health product recall
Quick summary

Quick summary: ARTIS Icono Ceiling, Floor and Biplane Systems. Reason: Health product recall. Category: Health products. Date: 2026-06-03.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: ARTIS Icono Ceiling, Floor and Biplane Systems. Reason: Health product recall. Category: Health products. Date: 2026-06-03.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandSiemens Healthcare Gmbh
ProductARTIS Icono Ceiling, Floor and Biplane Systems
IssueMedical devices - Performance
Recall classType II
IdentificationRA-82159

Affected products

ProductARTIS Icono Ceiling
LotSoftware version VE40A & VE40B
Model11328100
Serial numberSoftware version VE40A & VE40B
DescriptionARTIS Icono Ceiling
ProductARTIS Icono Floor
LotSoftware version VE40A & VE40B
Model11327700
Serial numberSoftware version VE40A & VE40B
DescriptionARTIS Icono Floor
ProductARTIS Icono Biplane
LotSoftware version VE40A & VE40B
Model11327600
Serial numberSoftware version VE40A & VE40B
DescriptionARTIS Icono Biplane

Issue

In very rare situations, after increasing the detector lift height to its maximum or minimum, the system may lose its movement capability. A reboot may not reestablish movement control. This failure can be resolved only by qualified service personnel.

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceGeneral public Healthcare Industry
CategoryHealth products - Medical devices - Radiology
CompaniesSiemens Healthcare GmbH Henkestr. 127, Erlangen, Germany, 91052
Recall date2026-05-21
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-82159
Background{}
CompaniesSiemens Healthcare GmbH Henkestr. 127, Erlangen, Germany, 91052
AudiencesGeneral public Healthcare Industry
Source updated2026-06-03

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.

Starting...

Related recalls

Similar recalls and safety notices.

Loading related recalls...