Aquilex Fluid Control System and Aquilex Fluid Control System Pump Assembly

2024-09-03

medium
Recall Safety notice
Quick summary

Quick summary: Aquilex Fluid Control System and Aquilex Fluid Control System Pump Assembly. Date: 2024-09-03.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Aquilex Fluid Control System and Aquilex Fluid Control System Pump Assembly. Date: 2024-09-03.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandW.O.M. World Of Medicine GmbH
ProductAquilex Fluid Control System and Aquilex Fluid Control System Pump Assembly
IssueMedical devices - Performance
Recall classType II
IdentificationRA-75981

Affected products

ProductAquilex Fluid Control System
LotMore than 100 numbers, contact manufacturer.
ModelAQL-100PBS RM-AQL-100PBS AQL-100P RM-AQL-100P
Serial numberMore than 100 numbers, contact manufacturer.
DescriptionAquilex Fluid Control System
ProductAquilex Fluid Control System Pump Assembly
LotMore than 100 numbers, contact manufacturer.
ModelAQL-100P
Serial numberMore than 100 numbers, contact manufacturer.
DescriptionAquilex Fluid Control System Pump Assembly

Issue

The Aquilex Fluid Control System is intended to provide fluid distension of the uterus during diagnostic and operative hysteroscopies and to monitor the volume differential between the irrigation fluid flowing into and out of the uterus. W.O.M. had received several complaints about deficit issues in the past. The manufacturer has observed that Aquilex Fluid Management Systems were used even though the system was mechanically damaged, or the deficit function was out of specification due to overdue preventive maintenance of the device. Failure investigation performed on complaint units indicates that the function test that is stipulated in the IFU is not always performed. The IFU stipulates that a function test must be performed before the system is used to detect a malfunction of the device at an early stage. The users should become more aware and perform the function tests, which detects any kind of deviation of the deficit measurement function. Therefore, W.O.M. will perform a product field action to add a 1000 g weight to the packaging of the pump and add the new labelling during biannually preventive maintenance. After preventive maintenance an information letter is enclosed with the device which explains the changes to labeling and the execution of the function test. Recall Start Date: May 28, 2021 Recall End Date: June 21, 2024 Old recall initiated in 2021 and web-posted delayed in September 2024 (closing date: 2024-06-21)

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - General hospital and personal use
CompaniesW.O.M. World Of Medicine Gmbh Salzufer 8, Berlin, Berlin, Germany, 10587
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-75981
Background{}
CompaniesW.O.M. World Of Medicine Gmbh Salzufer 8, Berlin, Berlin, Germany, 10587
AudiencesHealthcare
Source updated2024-09-03