Quick summary: Aquilex Fluid Control System. Date: 2023-08-15.
Aquilex Fluid Control System
2023-08-15
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Aquilex Fluid Control System. Date: 2023-08-15.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
WOM has received complaints from customers in the United States that the display of inflow volume on the Aquilex Fluid Control System can reach its limit of 30,000 ml during long procedures. In these cases, the display does not indicate the actual inflow volume. In cases where an unusually high amount of fluid is used for distension of the patient's uterus, the calculation of the inflow and outflow volumes reaches its limit and the pump control unit display will freeze at the maximum value. There is a risk of distension fluid reaching the circulatory system of the patient's soft tissue. Recall start date: July 17, 2023
What you should do
- Contact the manufacture if you require additional information.