Public recall detail

ApexPro Carescape Telemetry Server

2026-05-15

medium
Cardiovascular - General hospital and personal use Performance
Quick summary

Quick summary: ApexPro Carescape Telemetry Server. Reason: Performance. Category: Cardiovascular - General hospital and personal use. Date: 2026-05-15.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: ApexPro Carescape Telemetry Server. Reason: Performance. Category: Cardiovascular - General hospital and personal use. Date: 2026-05-15.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandGe Medical Systems Information Technologies Inc.
ProductApexPro Carescape Telemetry Server
IssueMedical devices - Performance
Recall classType II
IdentificationRA-82083

Affected products

ProductCarescape Telemetry Server Ark-2250l
LotMore than 10 numbers, contact manufacturer.
Model5697603
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionCarescape Telemetry Server Ark-2250l
ProductApexpro Host Software Ark2250 V6.0
LotMore than 10 numbers, contact manufacturer.
Model5878202
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionApexpro Host Software Ark2250 V6.0

Issue

Potential loss of ECG and SPo2 monitoring involving ApexPro Carescape Telemetry Server V5.0 and V6.0 can result from a persistent no comm or off network condition at the Carescape Central Station or CIC Pro Clinical Information Center.

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - Cardiovascular Health products - Medical devices - General hospital and personal use
CompaniesGe Medical Systems Information Technologies Inc. 3114 N Grandview Blvd, Waukesha, Wisconsin, United States, 53188
Recall date2026-05-06
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-82083
Background{}
CompaniesGe Medical Systems Information Technologies Inc. 3114 N Grandview Blvd, Waukesha, Wisconsin, United States, 53188
AudiencesHealthcare
Source updated2026-05-15

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.