Public recall detail

Anti-Lua (Lu1)

2022-04-07

medium
Recall Safety notice
Quick summary

Quick summary: Anti-Lua (Lu1). Date: 2022-04-07.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Anti-Lua (Lu1). Date: 2022-04-07.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandBio-Rad Medical Diagnostics Gmbh
ProductAnti-Lua (Lu1)
IssueMedical devices - Inaccurate test or measurement results
Recall classType II
IdentificationRA-64043

Affected products

ProductAnti-Lua (Lu1)
Lot3131020-00 3113160-00 3131030-00
Model808 216
Serial number3131020-00 3113160-00 3131030-00
DescriptionAnti-Lua (Lu1)

Issue

A recall was issued in connection to false negative results for the Anti-Lua control (Lu 1) from lot 3113160-00, because minimum antibody levels were not met. The two lots (3131020-00 and 3131030-00) produced to replace lot 3113160-00 have shown a decrease in reactivity over their life of approximately nine months. Soon, the acceptance criteria may no longer be met. Therefore, these lots are also being recalled. Recall start date: Mar 22, 2022

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceGeneral public Health professionals Hospitals
CategoryHealth products - Medical devices - Haematology
CompaniesBio-Rad Medical Diagnostics Gmbh Industriestr.1, Dreieich Germany, 63303
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-64043
Background{}
CompaniesBio-Rad Medical Diagnostics Gmbh Industriestr.1, Dreieich Germany, 63303
AudiencesGeneral public Health professionals Hospitals
Source updated2022-07-11

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.

Starting...

Related recalls

Similar recalls and safety notices.