Public recall detail

Allura Xper Systems

2025-08-15

medium
Recall Safety notice
Quick summary

Quick summary: Allura Xper Systems. Date: 2025-08-15.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: Allura Xper Systems. Date: 2025-08-15.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandPHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
ProductAllura Xper Systems
IssueMedical devices - Device compatibility
Recall classType II
IdentificationRA-77920

Affected products

ProductAllura Xper FD20/20 (Product Of Alluraclarity Family)
LotNot applicable.
Model722038
Serial numberNot applicable.
DescriptionAllura Xper FD20/20 (Product Of Alluraclarity Family)
ProductAllura Xper FD20 OR Table (Product Of Alluraclarity Family)
LotNot applicable.
Model722058
Serial numberNot applicable.
DescriptionAllura Xper FD20 OR Table (Product Of Alluraclarity Family)
ProductAllura Xper FD10
Lot608
Model001443
Serial number608
DescriptionAllura Xper FD10
ProductAllura Xper FD10/10 (Product Of Alluraclarity Family)
LotNot applicable.
Model722027
Serial numberNot applicable.
DescriptionAllura Xper FD10/10 (Product Of Alluraclarity Family)
ProductAllura Xper FD20/10 (Product Of Alluraclarity Family)
LotNot applicable.
Model722029
Serial numberNot applicable.
DescriptionAllura Xper FD20/10 (Product Of Alluraclarity Family)
ProductAllura Xper FD10 (Product Of Alluraclarity Family)
Lot1115
Model722026
Serial number1115
DescriptionAllura Xper FD10 (Product Of Alluraclarity Family)
ProductAllura Xper FD20 (Product Of Alluraclarity Family)
Lot1094
Model722028
Serial number1094
DescriptionAllura Xper FD20 (Product Of Alluraclarity Family)
ProductAllura Xper FD20 (Product Of Alluraclarity Family)
Lot1681
Model722028
Serial number1681
DescriptionAllura Xper FD20 (Product Of Alluraclarity Family)
ProductAllura Xper FD20 (Product Of Alluraclarity Family)
Lot2282
Model722028
Serial number2282
DescriptionAllura Xper FD20 (Product Of Alluraclarity Family)

Issue

Philips has identified that a component (air baffle) may be potentially missing in some Nehalem FlexVision-2 PCs that are part of the Allura XPER series. The air baffle directs part of the airflow inside the PC. When it is not installed, the temperature of the frame grabber card of the FlexVision PC-2 may be outside the specified temperature range, which may lead to an increased probability of failure of the frame grabber card. If this occurs, no image will be shown in the FlexVision monitor of the exam room. Recall start date: August 1, 2025

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceGeneral public Healthcare Industry
CategoryHealth products - Medical devices - Radiology
CompaniesPhilips Medical Systems Nederland B.V. Veenpluis 6, Best, Netherlands, 5684 PC
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-77920
Background{}
CompaniesPhilips Medical Systems Nederland B.V. Veenpluis 6, Best, Netherlands, 5684 PC
AudiencesGeneral public Healthcare Industry
Source updated2025-08-15

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.